Sondes ATR standard
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic ATR-Probes | |
Total Length | 1.5m (opt.:1m to 5m)* |
Shaft Length | 300mm (opt.: 100-700 mm)* |
Shaft Diameter | 12mm, 6.3mm |
Shaft Material | Hastelloy C22 |
Length of Legs | 500mm (opt. 200 – 500mm) |
Protective Tube Material | Liquid Tight SS-Conduit, KOPEX-Tube |
Minimal Bending Radius | 130mm |
Input / Output Connectors | SMA-905 (opt.: any other type) |
*Customized dimensions are available on request |
Specification of Fiber Optic Immersion ATR- Probes | |||||||
Probe type | Di- ATR | Ge- ATR | Si- ATR | ZrO₂- ATR | Sapphire-ATR | ||
Transmission range | 600-1900 cm-1 | 600-3100 cm-1 | 600-3100 cm-1 | 1550-9000 cm-1 | 400-2200 nm | 400-2200 nm | 180-1300 nm |
Fiber type | Silver Halide | Silver Halide | Silver Halide | Chalcogenide glass | Silica NIR | Silica NIR | Silica UV-VIS |
Temperature range | -150°C / + 140°C | -150°C / + 80°C | -150°C / + 140°C | -150°C / + 90°C | -150°C / + 200°C | -150°C / + 200°C | -150°C / + 200°C |
Pressure (max) | 200 Bar (300 Bar on request) | 10 Bar | 100 Bar | 100 Bar | 100 Bar | 100 Bar | 100 Bar |
Crystal-Shaft combination of ATR- Probes | ||||||
---|---|---|---|---|---|---|
Probe type | Di- ATR | Ge- ATR | Si- ATR | ZrO₂- ATR | Sapphire-ATR | |
400-2200nm | 1550 – 9000nm | |||||
Crystal -Schaft combination |
Sondes ATR à fibre optique pour les environnements difficiles
Les sondes ATR à fibre optique pour les environnements difficiles peuvent être utilisées pour la spectroscopie de process dans la gamme des IR proches et MIR pour le suivi de réactions dans une large gamme de températures allant de -150° à +250°C. Elles peuvent résister à des pressions élevées jusqu’à 200 bars et être utilisées avec des spectromètres FTIR ou tout autre spectromètre IR mais aussi avec des détecteurs spectraux dans le cadre d’un contrôle de processus automatisé .
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic High Temperature (HT) – ATR-Probes | |
---|---|
Total Length | 1.5m (opt.: 1m to 5m)* |
Shaft Length | 300mm (opt.: 300-700mm)* |
Shaft Diameter | 6.3mm or 12mm |
Shaft Material | Hastelloy C22 |
Length of Legs | 500mm (opt.: 200mm to 500mm) |
Protective Tube Material | Liquid Tight SS-Conduit, KOPEX-Tube |
Minimal Bending Radius | 130mm |
Input / Output Connectors | Long SMA (opt.: any other type) |
Cooling Air Flow Parameters | Excess pressure 0.5Bar, Flow 2300l/h |
Inner temperature control | Inside shaft to control ATR tip temperature |
Specification of Fiber Optic High Temperature (HT) – ATR- Probes | ||||
---|---|---|---|---|
Di-ATR | Si- ATR | ZrO₂- ATR | ZnSe-ATR | |
Probe type | ||||
Transmission range | 5.2-17μm600-1900 cm-1 | 3.2-17μm600-3100 cm-1 | 1.1-6.5μm1550-9000 cm-1 | 3.2-17μm600-3100 cm-1 |
Fiber type | PIR-900/1000Silver Halide | PIR-900/1000Silver Halide | CIR-500/550Chalcogenide glass | PIR-900/1000Silver Halide |
Temperature range | -150°C / +250°C | -150°C / +250°C | -150°C / +200°C | -150°C / +250°C |
Pressure (max) | 200 Bar | 100 Bar | 100 Bar | 10 Bar |
Sondes ATR à fibre optique pour applications en laboratoire
Nouvelles sondes ATR à fibre IR avec tige en PEEK économiques et parfaites pour une utilisation en laboratoire. Toutes les sondes ATR sont compatibles avec tous les types de spectromètres FT-IR et autres spectromètres IR, photomètres et détecteurs.
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic ATR-Probes for Lab applications | |
---|---|
Total Length | 1,5 m (opt: 1 to 5m) |
Shaft Length | 150mm (opt.:100 to 500mm) |
Shaft Diameter | 6.3mm |
Shaft Material | PEEK (polyetheretherketone) |
Length of Legs | 500mm (opt.: 200 to 500mm) |
Protective Tube Material | PEEK |
Minimum Bending Radius | 130mm |
Input / Output Connectors | Long SMA (opt.: any other type) |
Pressure (max) | 10 Bar |
Specification of Fiber Optic ATR-Probes for Lab Applications | ||||
---|---|---|---|---|
Probe type | ATR tip appearance | Transmission range | Fiber Type | Temperature range |
Diamond | 5.2-17μm 600 – 1900 cm-1 | PIR-900/1000 Silver Halide | -100°C / + 140°C | |
Ge- ATR | 3.2-17μm 600-3100 cm-1 | PIR-900/1000 Silver Halide | -100°C / + 80°C | |
ZnSe- ATR | 3.2-17μm 600-3100 cm-1 | PIR-900/1000 Silver Halide | -100°C / + 140°C | |
Si- ATR | 3.2-17μm 600-3100 cm-1 | PIR-900/1000 Silver Halide | -100°C / + 140°C | |
ZrO₂- ATR | 1.1-6.5μm 1550-9000 cm-1 | CIR 500/550 Chalcogenide glass | -100°C / + 90°C | |
Vis-NIR 400-2200cm-1 | Vis-NIR Silica | -100°C / + 140°C | ||
Sapphire-ATR | UV-Vis 190-1200cm-1 | UV-Vis Silica | -100°C / + 140°C | |
Vis-NIR 400-2200cm-1 | Vis-NIR Silica | -100°C / + 140°C |
Sondes ATR à fibre optique (conception Shaft-in-shaft)
Ces sondes ATR sont produites avec n’importe quel type d’élément ATR et pour n’importe quel type de spectromètre FTIR.
La conception unique imbriquée permet d’utiliser la sonde ATR dans les processus biologiques où la stérilisation est nécessaire sans avoir à renoncer aux avantages de la spectroscopie FTIR Mid-IR.
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic ATR-Probes | |
---|---|
Total Length | 1.5m (opt.:1m to 5m)* |
Shaft Length | 300mm (opt.: 100-700 mm)* |
Shaft Diameter | 12mm, 6.3mm (opt.:3mm) |
Shaft Material | Hastelloy C22 |
Length of Legs | 500mm (opt. 200 – 500mm) |
Protective Tube Material | Liquid Tight SS-Conduit, KOPEX-Tube |
Minimal Bending Radius | 130mm |
Input / Output Connectors | Long SMA (opt.: any other type) |
Compatible Process-Interfaces | Ceramat-FOS or SensoGate-FOS** |
*Customized dimensions are available on request **Available for 12mm shafts only |
Specification of Fiber Optic Immersion ATR- Probes | |||
---|---|---|---|
Probe type | Di- ATR | Si- ATR | ZrO₂- ATR |
Transmission range | 5.2-17μm 600-1900 cm-1 | 3.2-17μm 600-3100 cm-1 | 1.1-6.5μm 1550-9000 cm-1 |
Fiber type | PIR-900/1000 Silver Halide | PIR-900/1000 Silver Halide | CIR 500/550 Chalcogenide glass |
Temperature range | -100°C / + 140°C | -100°C / + 140°C | -100°C / + 90°C |
Pressure (max) | 200Bar (300 Bar) on request | 100 Bar | 100 Bar |
Sonde à boucle ATR à fibre optique
Les sondes à fibre PIR ATR-Loop sont bien adaptées à l’analyse à distance de la composition des liquides, des pâtes et des surfaces molles et ne nécessitent pas de préparation de l’échantillon.
Cette sonde à boucle est la plus simple et permet une spectroscopie ATR à faible coût.
L’utilisation de boucles avec différents nombres de spires (simple, double ou triple) permet de s’adapter à la sensibilité souhaitée.
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic Multi-Loop Probes FlexiSpec® | |
---|---|
Total Length | 1m (opt.: PIR up to 5m, CIR up to 10m) * |
Shaft Length | 120mm * |
Shaft Diameter | 10,6mm |
Shaft Material | PEEK |
Protective Tube Material | PEEK |
Input / Output Connectors | Long SMA-905 (opt.: any other type) |
Detachable PIR fiber Loop | Loop; Multi-Loop (Double-, Triple-Loop) |
*Customized dimensions are available on request |
Specification of Fiber Optic ATR-Loop Probes FlexiSpec® | ||
---|---|---|
Probe type | Chalcogenide-IR (CIR) | Polycrystalline-IR (PIR) |
Transmission range | 6500 – 1550 cm-1 (1.5-6.0μm) | 2500 – 600 cm -1 (4-17μm) |
Fiber type | Chalcogenide glass (As-S) | Silver Halide (AgCl:AgBr) |
Temperature range | ≤ 90°C | ≤ 100°C |
Sonde de contact ATR en diamant
Le « Touch Probe » ATR à extrémité plane en diamant est adapté aux mesures tactiles des solides et des fluides et se nettoie facilement grâce à la conception de la sonde.
La conception en cristal ATR à extrémité plate permet de mesurer au toucher des échantillons sensibles et des solides mous tels que la peau, la résine ou les polymères sans les endommager. En outre, les échantillons collants ou salissants comme les sirops, les colles ou les pâtes peuvent être facilement nettoyés de la sonde grâce à sa forme adaptée.
La comparaison des sondes ATR conique et plate démontre que la sonde à pointe plate permet à l’utilisateur d’obtenir une surface stable et constante avec des échantillons solides et mous. Cela garantit des résultats de mesure fiables et reproductibles, donnant à l’utilisateur des données quantitatives sur l’échantillon en question.
Points fort :
|
Applications:
|
Common Parameters of Fiber Optic ATR-Probes | |
---|---|
Probe Type | Diamond ATR |
Spectral Range | 5.2-17µm (600-1900cm-1) |
Fiber Type | PIR 900/1000 |
Tip Material | Diamond IIA, 2.5mm |
Sealing Material | Gold |
Shaft Material | Hastelloy C22 |
Minimal Bending Radius | 130mm |
Shaft Size | OD = 6.3mm, length = 300±5mm |
Flexible Fiber | 0.7m length + 2 legs of 0.5m each |
Total Length | shaft + fiber = 1.5±0.05m |
Termination | x 2 Long SMA |
Protection | Liquid Tight Protection (LTP) |